6) Specifications

General Specifications
Display :
3-3/4 digits 4000 counts LCD display

Polarity : Automatic

Update Rate : 3 per second nominal
Operating Temperature : 0

°C to 40°C

Relative Humidity : Maximum relative humidity 80% for temperature up to 31

°C

decreasing linearly to 50% relative humidity at 40

°C

Altitude: Operating below 2000m

Pollution degree: 2

Storage Temperature: -20

O

C to 60

O

C, < 80% R.H. (with battery removed)

Temperature Coefficient: nominal 0.15 x (specified accuracy)/

O

C @(0

O

C -18

O

C or

28

O

C -40

O

C), or otherwise specified

Sensing: Average sensing for AM-100. True RMS for AM-105 TRMS & AM-110 TRMS

Safety: The meter (all versions) is protected, against the users, by double insulation

per CSA C22.2 No. 1010-1-92, EN61010-1(1995) and IEC61010-1(1995) to CAT II
1000V & CAT III 600V.

Protection-reinforced CE Terminals (to COM) ratings:

V : Category II 1000 Volts AC & DC, and Category III 600 Volts AC & DC.
A / mA

µA : Category III 500 Volts AC and 300 Volts DC.

Overload Protections:

Protection-reinforced CE:
µA & mA : 0.63A/500V, IR200kA, F Fuse;
A: 12.5A/500V, IR20kA, F Fuse;

V: 1050Vrms, 1450Vpeak;

, & Others: 600VDC/VAC rms
Transient protection: 6.5kV (1.2/50

µs surge)

Power Supply: 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2

Power Consumption: 3.2 mA typical

Low Battery: Below approx. 2.4V

E.M.C.: Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and
EN61000-4-3(1996)

In an RF field of 3V/m:

Capacitance function is not specified

AC 4.000V range: Total Accuracy = Specified Accuracy + 700 digits
AC 400.0

µA range: Total Accuracy = Specified Accuracy + 300 digits

Other function ranges: Total Accuracy = Specified Accuracy + 40 digits
Performance above 3V/m is not specified