Amprobe AM-100 Digital-Multimeters User Manual

Page 13

Advertising
background image

12

Sleep Mode Timing: Idle for 30 minutes
Sleep Mode Consumption: 300

µA typical for AM100; 360µA typical for AM-105

TRMS & AM-110 TRMS

Dimension: L193mm X W97mm X H46
Weight: 370 gm

Special Features: 25ms Max Hold; Data Hold; Relative zero mode; Beep-jack™ input

warning; Back-lighted display (AM-110 TRMS only)

Accessories: Test leads (pair), batteries installed, user's manual

Electrical Specifications

Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23

O

C ±5

O

C & less than 75% R.H.

1)

Models AM-105 TRMS & AM-110 TRMS True RMS accuracy of ACV & ACA is

specified from 5 % (10% for AC400.0mV range) to 100 % of range, or otherwise

specified. Maximum Crest Factor < 1.75 : 1 at full scale & < 3.5 : 1 at half scale, and
with frequency components within the specified frequency bandwidth for non-sinusoidal

waveforms

DC Voltage
RANGE Accuracy
400.0 mV

0.3% + 4d

4.000V, 40.00V,
400.0V

0.5% + 3d

1000V

1.0% + 4d

NMRR: >50dB @ 50/60Hz
CMRR: >120dB @ DC, 50/60Hz, Rs=1k

Input Impedance: 10M

Ω, 30pF nominal

(1000M

Ω for 400.0mV range)

Type-K Temperature (AM-110 TRMS)
RANGE Accuracy*
-20

O

C TO 300

O

C 2%

+ 3

O

C

-4

O

F TO 572

O

F 2%

+ 6

O

F

*Type-K thermocouple range & accuracy

not included

AC Voltage
RANGE Accuracy

1)

50Hz -- 500Hz
400.0mV*

4.0% + 5d

4.000V, 40.00V,

400.0V

1.5% + 5d

1000V

4.0% + 5d

CMRR: >60dB @ DC to 60Hz, Rs=1k

Input Impedance: 10M

Ω, 30pF nominal

(1000M

Ω for 400.0mV range)

*Selection by RANGE button manually,
and is specified from AC 10mV (AC

40mV for True RMS models AM-105

TRMS AM-110 TRMS) and up


Max Hold (Voltage & Current)

Specified accuracy ± 50 digits for

changes > 25ms in duration

Advertising
This manual is related to the following products: