Amprobe AM-60 Compact-Digital-Multimeter User Manual

Page 15

Advertising
background image

10

* AC Conversion Type : Conversion type and additional specification are same
as DC/AC Voltage.

** Ampere Testing Duty Ratio Table

Ampere

Testing Time

Rest Time

10A

1min

10min

9A

2min

10min

8A

3min

10min

7A

4min

10min

6A

5min

10min

5A

Continually

N/A

Resistance

Range

Accuracy

Voltage Burden

200.0~200.0Ke **

±(0.7% + 3 dgt)

2V max

2.000Me **

±(1.0% + 3 dgt)

20.00Me *

±(1.5% + 3 dgt)

Open circuit Voltage:

-1.3V approx.

* < 100 dgt rolling.
** The minimum LCD reading is 1400 count in Auto Ranging Mode.

Diode Check and Continuity

Resolution

Accuracy

10 mV

±(1.5% + 5 dgt)*

* For 0.4V ~ 0.8V
Max.Test Current : 1.5mA
Max. Open Circuit Voltage : 2V
Overload Protection : 600V rms.