EMC: This product complies with requirements of the following European

Community Directives: 89/336/EEC (Electromagnetic Compatibility) as amended by
93/68/EEC (CE Marking). Directive73/23/EEC (Low Voltage) does not apply to this
product. However, electrical noise or intense electromagnetic fields in the vicinity
of the equipment may disturb the measurement circuit. Measuring instruments will
also respond to unwanted signals that may be present within the measurement
circuit. Users should exercise care and take appropriate precautions to avoid
misleading results when making measurements in the presence of electronic
interference.

ENVIRONMENTAL CONDITIONS:

1. Indoor Use.
2. Operating Ambient: 0% to 80% R.H. (0 to 35 ºC); 0% to 70% (35 to 50ºC) R.H.
3. Storage Temperature: -20 to 60ºC, 0 to 80% R.H. with battery removed from

Temperature Converter.

ELECTRICAL SPECIFICATIONS:

MEASUREMENT RANGE

ACCURACY

INPUT PROTECTION

-50ºC -20ºC

=(2.0% rdg + 2ºC)

-19ºC 350ºC

=(0.5% rdg + 2ºC)

351º C 500ºC

=(2.0% rdg + 2ºC)

501ºC 1000ºC

=(2.9% rdg + 2ºC)

-58ºF -4ºF

=(2.0% rdg + 4ºF)

-3ºF 662ºF

=(0.5% rdg + 4ºF)

663ºF 932ºF

=(2.0% rdg + 4ºF)

933ºF 1832ºF

=(2.9% rdg + 4ºF)

60Vdc & 24Vrms

ACCESSORIES:

TP-255A

Replacement Bead Probe

TP-254A

Immersion Probe with Handle

MH101KA Master Handle Probe
TP110KA

Immersion Probe *

TP120KA

Surface Probe *

TP130KA

Air Probe *

*(TP110KA, TP120KA, & TP130KA require MH101KA Handle)