PREPARATION AND CAUTION BEFORE MEASUREMENT

1. Before measurement, warm up for at least 30 seconds, after connecting the

thermocouple to the connector of the Temperature Converter.

2. If the instrument is used near noise generating equipment, be aware that the

Voltage output may become unstable or indicate large errors.

3. Check TC-253B battery. The multimeter should read > 7.1 volts in the OFF

position.

TEMPERATURE MEASUREMENTS

1. Warning! do not measure the temperature of any surface that the potential

exceeds 60 Vdc or 24 Vrms.

2. Connect the plug of the thermocouple to the connector of the thermometer.
3. Select the ºC/ºF function desired.
4. Insert the “Output Plugs (V and COM)” to the input terminals of Multimeter.

Set the Multimeter to “DC mV or V” range.

5. Place the sensing point of thermocouple on the surface or environment to be

measured.

6. Wait until the reading stabilizes.

GENERAL SPECIFICATIONS

This Temperature Converter conforms to the temperature/voltage tables of the
National Bureau of Standards and to the IEC 584 Standards for K-type.
1. Temperature Scale: Celsius or Fahrenheit user selectable.
2. Input: Single K-type thermocouple.
3. Output to Meter: 1mVdc per ºC or ºF.
4. Measurement Range: -50ºC to 1000ºC – (-58ºF to 1832ºF).
5. Low Battery Indication: if the reading is less than 7.1V.
6. Temperature Coefficient: 0.15 x (specified accuracy) / ºC, (<18ºC or >28ºC).
7. Power Requirements: Standard 9V battery, NADA 1604, JIS 006P, or IEC6F22 Size.
8. Battery Life: Alkaline 300 hours.
9. Dimensions: 122mm(L) x 46mm(W) x 30mm(D).
10. Weight (including battery): 114 grams.
11. Accessories: K-type bead thermocouple, battery (installed), operators