Specifications, 1 general information, 2 electrical specifications – Amprobe AMB-25 Insulation-Tester User Manual

Page 2

Advertising
background image

2-2 Electrical Specifications

2. SPECIFICATIONS

Accuracies are specified as:

±(...% of reading + ...digits) at 23℃ ±5℃,below 80% RH.

2-1 General Information

Environment conditions :

‰

Insulation Resistance

c

Installation Categories

Ⅲ 1000V

Range

Resolution

Accuracy

Test

Voltages

40MΩ/400MΩ

250V+20% ~ -0%

400MΩ/1000MΩ

/500V

500V+20% ~ -0%

400MΩ/ 2000MΩ

/1000V

40MΩ: 10 KΩ

400MΩ:100 KΩ

4000MΩ: 10MΩ

3%+5

(<1000M)

5%+5

(>1000M)

1000V+20% ~ -0%

Analog Bar Graph

0 to 4000MΩ

Nominal Current

≧ 1mA

d

Pollution Degree 2

e

Altitude up to 2000 meters

f

Indoor use only

Safety Meets of IEC61010-1 and IEC61557
Display : Dual display, 3-3/4 Digital readout with analog bar

indication.

Sampling Rate : 2.5 sample/sec.
Manual data Memory and Read : Memory capacity 9 set.

Over Range Indicator :

” will be displayed.

Low Battery Indication :

BT

‰

Resistance & Continuity (<40Ω)

The (

) will be displayed when the battery voltage drop

below the operating voltage.

Range

Resolution

Accuracy

Max. open Circuit

Voltage

Overload

Protection

999.9Ω

0.1Ω

1%+2

3V

1000Vrms

Operating Temperature and Humidity:

0℃ to 50℃ (32℉ to 122℉) below 80% RH (noncondensing).

Temperature Coefficient : 0.10 x (specified accuracy)/ ℃

Storage Temperature and Humidity :

-10℃ to 60℃ (14℉ to 140℉) below 70% RH (noncondensing)


Battery : 6 x 1.5V Size “AA” battery.

Fuse : 6mm x 32mm (0.25 x 1.25 inch), 0.5A 1000V, Fast Acting.

Dimensions : 235 (L) x 116 (W) x 54(H) mm ,

(9.3”L x 4.6”W x 2.1”H)



Weight : Approx. 520g (1.15 LB), include battery

Accessories : Test leads, 6pcs battery, Holster, operation manual.