Amprobe AMP-25 User Manual

Page 18

Advertising
background image

14

ELECTRICAL SPECIFICATIONS

Accuracy is given as ±(% of reading + counts of least
significant digit) at 23°C ± 5°C,with relative humidity less
than 80% R.H., AC A specifications are ac coupled, true
R.M.S. The crest factor may be up to 3.0 as 4000 counts.
For non-sinusoidal waveforms, additional accuracy by Crest
Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0
Add 5.0% for C.F. 2.0 ~ 2.5
Add 7.0% for C.F. 2.5 ~ 3.0
Position Error of Clamp: ±1.5% of display reading

AC Current

Range

60.00 A

300.0 A

Resolution

0.01 A

0.1 A

Accuracy

50 Hz to 100 Hz

± (1.5 % + 25 LSD)

at < 3 A

± (1.5 % + 5 LSD)

at ≥ 3 A

± (1.5 % + 5 LSD)

Accuracy

100 Hz to 400 Hz

± (2.5 % + 25 LSD)

at < 3 A

± (2.5 % + 5 LSD)

at ≥ 3 A

± (2.5 % + 5 LSD)

Frequency Response: 50 to 400Hz (Sine Wave)

DC Current

Range

60.00 A

1)

300.0 A

Resolution

0.01 A

0.1 A

Accuracy

± (1.5 % + 10 LSD)

2)

± (1.5 % + 5 LSD)

1)

There are less than 0.3A variations as measuring in

different directions.

2)

Add 10 LSD to accuracy in Auto AC / DC A mode.