HP 16500C User Manual

Page 20

Advertising
background image

To capture a write of known bad data to a particular
variable

The trigger specification ANDs the bad data on the data bus, write
transaction on the status bus, and address of the variable on the address bus.

1

Select the state analyzer Trigger menu.

2

Define the terms BAD_DATA, WRITE, and VAR_ADDR to represent
the bad data value, write status, and the address of the variable.

3

Under State Sequence Levels, enter the following sequence
specification:

While storing “anystate” TRIGGER on “BAD_DATA

WRITE

VAR_ADDR” one time (you use the Combination trigger term to do this)

Store “anystate”

Capturing a Bad Write to a Variable

Triggering
To capture a write of known bad data to a particular variable

1–10