pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 14 / 67

pH ARC sensor: 4-20mA interface - linear

-15

-10

-5

0

5

10

15

20

25

30

35

0

4

8

12

16

20

24

Current [mA]

te

m

p

er

at

u

re

C

]

1

2

3

4

5

6

7

8

9

10

11

p

H

Temp [°C]

pH

Figure 2.4.5.6.3: Example of linear 4-20 mA output characteristics for pH or temperature.

Current

pH

Temperature

4 mA

2

-10

°C

20 mA

9

+30

°C

pH ARC sensor: 4-20mA interface - bilinear

-15

-10

-5

0

5

10

15

20

25

30

35

0

4

8

12

16

20

24

Current [mA]

te

m

p

e

ra

tu

re

C

]

1

2

3

4

5

6

7

8

9

10

11

p

H

Temp [°C]

pH

Figure 2.4.5.6.4: Example of bilinear 4-20 mA output characteristics for pH or temperature.

Current

pH

Temperature

4 mA

2

-10

°C

12 mA

9

+20

°C

20 mA

10

+30

°C