2.6.3 PA12: Moving Average R

pH ARC Sensors allow to have a separate moving average on secondary measurement values:

 Glass resistance

 Reference resistance

The moving average can be applied on 1 to 16 3-s measurement values. The default value is 4.
Especially if high resistances are measured, it is recommended to choose a higher moving average.

2.6.3.1

Description of PA12 (Moving Average R)

In register 3456, a plain text ASCII description of PA12 is given.

Start
register

Number of
registers

Reg1

… Reg8

16 ASCII characters

Modbus
function code

Read
access

Write
access

3456

8

Description of PA12

3, 4

U/A/S

none

Figure 2.6.3.1.1: Definition of register 3456.

Command: Moving average text

Modbus address:

3456

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

Moving average R

Figure 2.6.3.1.2: Example to read the description for

“

Moving average R

“

.

2.6.3.2

Selecting the Physical Unit and Writing the Value for PA12

In register 3464, the available physical units for PA12 are defined.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

3464

2

Available physical units for PA12

3, 4

U/A/S

none

Figure 2.6.3.2.1: Definition of register 3368.

Command: Moving average av. units

Modbus address:

3464

Length:

2

Type:

3

Read

Parameter:

Units

Format:

Hex

Value:

0x01

Figure 2.6.3.2.2: Example to read the available physical units for PA12. The only one available here is

“

none

”(0x01). For the definition of the physical units see chapter 2.5.1.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

3466

4

Select physical unit
for PA12

Value for PA12
(1-16, default: 4)

16

none

S

Figure 2.6.3.2.3: Definition of register 3466. Only one bit for the physical unit can be set. PA12 can be

set to the values 1-16.