5 reading definition of sip and cip – Hamilton Electronics EPHUM011 User Manual

Page 62

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 62 / 67

2.8.5 Reading Definition of SIP and CIP

pH ARC Sensor are counting special cleaning events such as sterilizations or cleaning cycles by
means of tracking typical temperature profiles (see chapter 2.8.2).

Register 4988 defines a typical temperature profile for SIP (sterilization in place) and register 4996 for
CIP (cleaning in place). For the explanation the following values are given:

CIP temperature min: 80 °C

CIP temperature max: 100 °C

CIP time min: 30 minutes

SIP temperature min: 120 °C

SIP temperature max: 130 °C

SIP time min: 30 minutes

CIP and SIP

0

20

40

60

80

100

120

140

0

8

:0

0

0

8

:1

5

0

8

:3

0

0

8

:4

5

0

9

:0

0

0

9

:1

5

0

9

:3

0

0

9

:4

5

1

0

:0

0

1

0

:1

5

1

0

:3

0

1

0

:4

5

1

1

:0

0

1

1

:1

5

1

1

:3

0

1

1

:4

5

1

2

:0

0

1

2

:1

5

1

2

:3

0

1

2

:4

5

1

3

:0

0

1

3

:1

5

1

3

:3

0

1

3

:4

5

1

4

:0

0

1

4

:1

5

1

4

:3

0

1

4

:4

5

1

5

:0

0

1

5

:1

5

1

5

:3

0

1

5

:4

5

1

6

:0

0

1

6

:1

5

1

6

:3

0

1

6

:4

5

1

7

:0

0

Time

T

em

p

er

at

u

re

C

]

T CIP

T SIP

CIP min

CIP max

SIP min

SIP max

no SIP 1)

SIP ok 3)

CIP ok 2)

no CIP 4)

no CIP 4)

Figure 2.8.5.1: Definition of CIP and SIP cycles.

1) no SIP-cycle counted, because time too short <30 minutes.
2) CIP-cycle counted, because time >30 minutes and in CIP temperature range.
3) SIP-cycle counted, because time >30 minutes and in SIP temperature range.
4) no CIP-cycle counted, because of reaching the SIP-min limit.

Start
register

Number
of
registers

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

4988

8

SIP
Tempera-
ture min
[°C]

SIP
Tempera-
ture max
[°C]

SIP
Process
time min
[min]

Empty

3, 4

U/A/S

S

4996

8

CIP
Tempera-
ture min
[°C]

CIP
Tempera-
-ture max
[°C]

CIP
Process
time min
[min]

Empty

3, 4

U/A/S

S

Figure 2.8.5.2: Definition of register 4988 and 4996.