HP E1368A User Manual

Page 45

Advertising
background image

For the example, use:

an HP-IB select code of 7, primary address of 09, and secondary
address of 15 for the Microwave Switch

an HP Series 200/300 Computer with HP BASIC

Execute:

10

!Call computer subprogram “Errmsg” if a Microwave

20

!Switch programming error occurs. Enable the computer to respond

30

!to an interrupt from the Microwave Switch.

35

!

40

ON INTR 7 CALL Errmsg

50

ENABLE INTR 7:2

60

!

70

!Unmask the Event Status bit in the Microwave Switch’s Status

80

!Register. Unmask the Microwave Switch error conditions in its

90

!Standard Event Status Register.

95

!

100 OUTPUT 70915;"*SRE 32"

110 OUTPUT 70915;"*ESE 60"

120 !

130 !At this point, the Microwave Switch is programmed for the intended

140 !application.

145 !

150 OUTPUT 70915;" . . .

160 OUTPUT 70915;" . . .

170 ENTER 70915; . . .

180 PRINT . . .

190 END

200 !

210 !When an error occurs, clear the Microwave Switch to regain

220 !control. Execute a Serial Poll to clear the service request

230 !bit in the Status Register. Read all error messages in the

240 !Microwave Switch error queue. Clear all bits in the counter

250 !Standard Event Status Register.

255 !

260 SUB Errmsg

270 DIM Message$[256]

280 CLEAR 70915

290 B=SPOLL (70915

300 REPEAT

310 OUTPUT 70915; “SYST:ERR?”

320 ENTER 70915; Code, Message$

330 PRINT Code,Message$

340 UNTIL Code=0

44 Using the HP E1368A/69A/70A Modules

Chapter 3

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com