Writing to the registers, Status/control register, Channel enable register – HP E1368A User Manual

Page 78

Advertising
background image

Status/Control

Register

Each coaxial switch requires about 30 msec execution time (close to open or
open to close) during which time the switch is "busy". A read of the
Status/Control Register (base + 04

16

) returns FFFF

16

when the module is

not busy or returns FF7F

16

(bit 7 is low-true) when the module is busy.

Channel Enable

Register

A read of the Channel Register (base + 08

16

) always returns FFFF

16

,

regardless of the channel states.

Writing to the Registers

You can write the following Microwave Switch registers:

Channel Enable Register (base + 08

16

)

Channel Enable

Register

Writes to the Channel Enable Register (base + 08

16

) enable you to open or

close the desired channel. For example, write a "1" to bit 2 of the Channel
Enable Register to close channel 02. Or, write a "0" to bit of the register to
open channel 00.

Note

All HP 3331X and HP 876X Microwave Switches are latching and will
remain in the last state during power down. Performing a reset (or at
power-up) opens all channels (refer to Figure 2-6). Any combination of
channel open or closed states is allowed.

Appendix B

HP E1368A/69A/70A Registers 77