Parameter Types. The following table contains explanations and examples
of parameter types you might see later in this chapter.

Parameter Type

Explanations and Examples

Numeric

Accepts all commonly used decimal representations of
numbers including optional signs, decimal points, and
scientific notation.

123, 123E2, -123, -1.23E2, .123, 1.23E-2, 1.23000E-01.
Special cases include MIN, MAX, and DEF. MIN selects
minimum value available, MAX selects maximum value
available, and DEF selects default or reset value.

Boolean

Represents a single binary condition that is either true or
false.

1 or ON; 0 or OFF.

Discrete

Selects from a finite number of values. These parameters
use mnemonics to represent each valid setting.

An example is the TRIGger:SOURce <

source

> command

where

source

can be BUS, EXT, HOLD, or IMM.

Optional Parameters. Parameters shown within square brackets ([ ]) are
optional parameters. (Note that the brackets are not part of the command
and are not sent to the instrument.) If you do not specify a value for an
optional parameter, the instrument chooses a default value. For example,
consider the

ARM:COUNt? [<MIN | MAX>]

command. If you send the

command without specifying a parameter, the present

ARM:COUNt

value is

returned. If you send the

MIN

parameter, the command returns the

minimum count available (1). If you send the

MAX

parameter, the

command returns the maximum count available (32,767). Be sure to place a
space between the command and the parameter.

Linking Commands

Linking IEEE 488.2 Common Commands with SCPI Commands. Use
a semicolon between the commands. For example:

*RST;OUTP ON

or

TRIG:SOUR

HOLD;*TRG

Linking Multiple SCPI Commands. Use both a semicolon and a colon
between the commands. For example:

ARM COUN 1;:TRIG:SOUR EXT

SCPI allows several commands within the same subsystem to be linked
with a semicolon. For example:

ROUT:CLOS (@100);:ROUT:OPEN (A101)

or

ROUT:CLOS (@100);OPEN (@101)

Chapter 4

HP E1368A/69A/70A Command Reference 49