Input n:clock[:source – HP E1459A User Manual

Page 48

Advertising
background image

HP E1459A SCPI Command Reference

47

INPut

n:CLOCk[:SOURce]?

Returns the programmed value of the input clock source for Port n.

Parameters

Comments

•

The value returned is an unquoted string of

EXT

or

INT

.

Example

INPut2:CLOCk[:SOURce]?

queries the input clock source for Port 2.

INPut

n:DEBounce:TIME <time> | MINimum | MAXimum | DEFault

Programs the input debounce time for Port n. Ports 0 and 1 share the same debounce
circuitry; Ports 2 and 3 share the same debounce circuitry.

Parameters

Comments

•

The available debounce times are:

18.0

µ

S, 36

µ

S, 72

µ

S, 144

µ

S, 288

µ

S, 576

µ

S, 1.13 mS, 2.26 mS, 4.6 mS, 9.2

mS, 18.4 mS, 36.9 mS, 73.8 mS, 148.0 mS, 294.0 mS, 590.0 mS, 1.18 S, 2.36
S, 4.72 S, 9.43 S, 18.9 S, 37.8 S, 75.0 S, 150.0 S, 300 S, 600 S, 1200 S, 2400 S,
4800 S, 9600 S. Any value sent other than those listed is rounded to the nearest
discrete value.

•

Debounce time is rounded to the nearest discrete value. For example 16.0

µ

S to

18.4

µ

S is rounded to 18

µ

S, 18.5

µ

S to 36.4

µ

S is rounded to 36

µ

S, 36.5

µ

S

to 72.4

µ

S is rounded to 72

µ

S, etc.

•

Ports 0 and 1 use the same debounce time, Ports 2 and 3 use the same debounce

time. For n = 0 or n = 1, both Ports 0 and 1 are set; for n = 2 or n = 3, both Ports
2 and 3 are set.

•

*RST Condition: sets the value for the debounce time to 18.0

µ

Sec for all four

Ports.

Parameter Name

Parameter Type

Range of Values

Default

INPut

n

numeric

0, 1, 2, 3

0

Parameter Name

Parameter Type

Range of Values

Default

INPut

n

<time>

DEFault

MINimum

MAXimum

numeric

numeric (floating pt)

0, 1, 2, 3

18.0

µ

Sec through 9600 Sec

Default 18.0

µ

Sec

Minimum 18.0

µ

Sec

Maximum 9600 Sec

0

18.0

µ

Sec

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com