Fluke Ti20 User Manual

Page 44

Advertising
background image

Ti20

Users Manual

3-6

Adjusting Reflected Temperature Compensation Values

As the RTC value changes, the live thermal image will also change as the displayed

temperature changes. The default RTC setting is 100

°

C (212

°

F)and the adjustment

range is from -50

°

C to 600

°

C (-58

°

F to 1112

°

F). RTC is OFF in the factory default

state.

1. From the Home Display, press G (

MENU

) three times to navigate to the Adjust

Emissivity and RTC display.

2. Press F (

RTC

) to access the Adjust RTC display.

3. Press F (

RTC

) to toggle RTC OFF/ON or press H (

ADJUST

) to access the

Adjust RTC display.