Accuracy – Fluke RUSKA 2470 User Manual

Page 12

Advertising
background image

RUSKA 2470

Users Manual

1-4

Temperature Range

Operating

18 ºC to 28 ºC

Storage

-40 ºC to 70 ºC when thermometer and low range piston & Cylinder are
stored separately from each other.

Humidity Range

Operating

20% to 75% noncondensing

Storage

0% to 90% noncondensing

Pressure Medium

Clean dry gas, Nitrogen or equivalent, regulated to a pressure Compatible
with each particular Piston/cylinder assembly. Dew Point of less than or
equal to -60 ºF

Pressure

Maximum Working Pressure

2470 piston/Cylinder installed in 2470 Column adapter

With High Range Piston/Cylinder

3000 psig

2465 piston/Cylinder installed in 2465 Column adapter

With Mid Range Piston/Cylinder

100 psig (Do Not Exceed 6.31 Kg Mass Load)

With Low Range Piston/Cylinder

25 psig (Do Not Exceed 6.31 Kg Mass Load)

2468 piston/Cylinder installed in 2465 Column adapter

With Mid Range Piston/Cylinder

200 psig (Do Not Exceed 12.31 Kg Mass Load)

With Low Range Piston/Cylinder

50 psig (Do Not Exceed 12.31 Kg Mass Load)

Note

The pressurized housing with the 2470 adapter has been tested to 4000 psig.
The pressurized housing with the 2465 adapter has been tested to 1000 psig.

Accuracy

The gauge is capable of measuring pressures to the accuracy indicated below. See the

calibration report for the actual accuracy of this gauge. The claim for accuracy is valid

only when the gauge is operated according to the instructions provided with the

equipment. In addition, the claim is valid when the value of gravity acting upon the

weights is known to +/-0.001 cm sec

2

.

Accuracy is defined as the departure of the measured pressure from the true pressure.

The value is based on a simple error analysis of the calibration experiment and represents

the sum of the systematic errors and two standard deviations of the random variability of

the measurement process.

Gauge Pressure (0.0035% Class)

High Range Piston
(0.013 sq in area)

0.0035 percent of reading
or 0.002 psi, whichever is greater

Mid Range Piston
(0.13 sq in area)

0.0035 percent of reading
or 0.0002 psi, whichever is greater

Low Range Piston
(0.52 sq in area)

0.0035 percent of reading
or 0.00005 psi, whichever is greater