Fluke RUSKA 2470 User Manual

Page 28

Advertising
background image

RUSKA 2470

Users Manual

4-2

4. Mount the FPIs (Float Position Sensors) to the blocks, orient the FPI sensors such

that their cables are trail away from the base in the same direction as the PRT cable.

5. Route the cables from the FPIs and the PRT such that they do not interfere with other

activities on the calibration bench.

6. Connect the PRT and FPIs to the RUSKA 2456 per instructions found in the RUSKA

2456 Manual.

7. Readjust the leveling screws if necessary until gauge is level in 2 axes by using the

bubble gauge.

8. Install the desired piston and cylinder assembly into the housing. See Chapter 5,

Operation, for operation of the gauge.

9. If the optional glass thermometer is being used, insert it into the 1/4 NPT hole in the

right hand side of the Base Assembly. The graduated end of the thermometer will

protrude through the hole in the right hand side of the base casting. Here too, a small

amount of heat sink compound applied to the bulb of the thermometer will shorten

the response time of the thermometer.

W

Caution

Always remove the thermometer before shipping the gauge

base.

gmq06.bmp

Figure 4-1. FPI Mounting - RUSKA 2456 Piston Gauge Monitor

Advertising