Establishing pressure, Gauge pressures – Fluke RUSKA 2470 User Manual

Page 43

Advertising
background image

Operation

Establishing Pressure

5

5-13

gmq23.bmp

Figure 5-15. Order of Installation

Establishing Pressure

Gauge Pressures

Pressure should be admitted slowly into the gauge, preferably through a metering valve.

It is very convenient to have a volume adjusting pump such as RUSKA Series 3990

Manual Pressure Control Packs to adjust the system pressure. The RUSKA Series 3990

Manual Pressure Control Packs provides a supply gas regulator, valves, monitor gauge,

and system volume adjuster all integrated into a compact package. The RUSKA Series

3990 Manual Pressure Control Packs facilitates the control of the input pressure to the

gauge and provides facility to change the float position.
The index mark for the plane of reference on the RUSKA 2470 base is a circumferential

groove around the base of the column. This reference is compared to the bottom face of

the hanger mass as shown in Figure 5-16.