To recalibrate the test tool, refer to Figure 6-1 and proceed as follows:

Test Tool

Smart Remote

15 cm

UTP Patch Cable

gc26f.eps

Figure 6-1. Connections for Self-Calibration (DSP-2000 Shown)

1.

Turn the rotary switch to SPECIAL FUNCTIONS.

2.

Use

D to highlight Self Calibration.

3.

Press

E.

4.

On Model DSP-100, remove any cable connected to the test tool’s BNC
connector.

5.

Connect the test tool to the remote as directed by the displayed messages. On
Model DSP-2000, connect the cable to the CABLE TEST jack.

6.

Press

T to start the calibration.

When the calibration is complete, the calibration data and the remote’s serial
number are automatically stored in the test tool’s calibration library.

If the message

SELF CALIBRATION FAIL appears, check the following:

Verify that the main test tool and the remote are connected with the patch
cable described on the test tool’s display.

Check the patch cable for breaks or damaged connectors.

Check the connectors on the main and remote units for damage.

If the self calibration continues to fail, send both the main and remote units to the
nearest Fluke service center. See “Service Center Repair” in Chapter 8 for details.