Resistance, Length – Fluke DSP-2000 User Manual

Page 56

Advertising
background image

DSP-100/2000

Users Manual

3-8

Resistance

The resistance test measures the dc loop resistance for each cable pair. The
resistance results screen displays the resistance, limit, and pass/fail result for each
cable pair. A

PASS result means that the measured resistance is less than the

limit. A

FAIL result means that the measured resistance exceeds the limit.

Length

The length test measures the length of each cable pair tested. The main Autotest
results screen shows the length of the cable pair having the shortest electrical
delay. Length is displayed in meters or feet. The length results screen displays the
length, limit, and pass/fail result for each cable pair. You can change the length
units in the SETUP mode, as described in “Selecting a Length Unit” in Chapter 2.

Notes

A 2 to 5 percent difference in measured length between twisted pairs
is typical. This difference is due to differences in the number of
twists in the cable pairs.

Differences between measured and actual values of cable length can
be caused by variations in the cable’s NVP value. To ensure
maximum accuracy of length measurements, perform an NVP
calibration as described in Chapter 6.

The length test limits include an extra 10% to account for variations
in NVP.

A

PASS result means that the measured length is within the specified limit for the

selected test standard. A

FAIL result means that the measured length exceeds the

limit.