Attenuation – Fluke DSP-2000 User Manual

Page 58

Advertising
background image

DSP-100/2000

Users Manual

3-10

Attenuation

Note

Incorrect conduit or temperature settings can cause false
attenuation results. You can change these settings in the SETUP
mode, as described in “Configuring the Test Tool” in Chapter 2.

The attenuation test measures the loss of signal strength over the length of the
cable.

The first attenuation results screen shows the cable pairs tested, the worst-case
attenuation margin found, and a

PASS or FAIL result for each pair.

To see detailed results for the cable pairs, use

D U to highlight a cable pair, then

press

@ View Result. Table 3-2 describes the items on the attenuation

results screen.

Table 3-2. Items on the Attenuation Results Screen

Item

Description

Pair

The cable pair relevant to the results.

Result

The overall result for the test. A

PASS

result means that measured

attenuation is lower than the specified limit for the selected test standard. A

FAIL

result means that the measured attenuation is higher than the

specified limit.

Attenuation

If the test passed, this value is the highest measured attenuation. If the test
failed, this value is the highest measured attenuation that exceeds the test
limits.

Frequency

If the test passed, this frequency is where the highest measured attenuation
occurred. If the test failed, this is where the highest failing value of
attenuation occurred.

Limit

The highest attenuation value acceptable at the frequency shown. This value
is based on the maximum allowable cable length.

Margin

The difference between the worst-case attenuation and the limit. A positive
number means that the measured attenuation value is lower than the limit. A
negative number means that the attenuation is higher than the limit.

Advertising
This manual is related to the following products: