Setting up the beeper, Viewing instrument information, Viewing memory – Fluke ESA612 User Manual

Page 26: Setting the gfci limit, Performing electrical safety tests

Advertising
background image

ESA612
Users Manual

12

1. From the setup menu, press the softkey labeled

More twice to set F1 to the display contrast function.

2. Press the softkey labeled Display Contrast.
3. Press

G

or H to increase or decrease the display’s

contrast respectively.

4. Press the softkey labeled Done to exit contrast

setup.

Setting up the Beeper

To enable or disable the beeper:
1. From the setup menu, press the softkey labeled

More twice to set F2 to the beeper on/off function.

2. Press the softkey labeled Beeper to switch the

beeper on and off.

3. Press the softkey labeled Done to go back to the

setup menu.

Viewing Instrument Information

To view information about the Analyzer:
1. From the setup menu, press the softkey labeled

More twice to set F3 to the instrument information
function.

2. Press the softkey labeled Instrument Information.
3. After viewing the displayed information, press the

softkey labeled Done to exit the information screen.

Viewing Memory

Refer to the Using Memory section later in the manual to
learn about viewing memory and how to store data in the
Analyzer.

Setting the GFCI Limit

To set the GFCI current limit:
1. From the setup menu, press the softkey labeled

More to reveal additional menu selections.

2. Press the softkey labeled Instrument to reveal the

instrument setup selections.

3. Press the softkey labeled GFCI Limit to open the

scroll box above the softkey label.

4. Press

G

or H to adjust the current limit from 5 to

25 mA.

5. Press the softkey labeled Done to exit the GFCI Limit

setup function.

Performing Electrical Safety Tests

The Analyzer is designed to perform a number of different
electrical and performance tests on biomedical
equipment. The following sections describe the various
tests and how to perform them using the Analyzer.