Performing a Differential Leakage Current Test

Note

The Differential Leakage Current test is available
when the EN62353 standard is selected.

The differential leakage current test measures the
magnitudes of the differential current flowing in the
Equipment Outlet live and neutral, with power applied to
the equipment outlet. All applied parts should be
connected during this test, if equipment has applicable
applied parts.
To perform a differential leakage current test:
1. Press

B

.

2. Press the soft key labeled Differential.
Figure 28 shows the electrical connections between the
Analyzer and the DUT during a Differential Leakage
Current test.
The following outlet conditions apply when performing this
test:

Normal Polarity, Closed Earth

Normal Polarity, Open Earth

Reversed Polarity, Closed Earth

Reversed Polarity, Open Earth

Note

If there are more than five applied parts to
connect to the Analyzer, see Using the 1210
Adapter later in this manual.