Fluke ESA612 User Manual

Page 77

Advertising
background image

Electrical Safety Analyzer

Detailed Specifications

63

Ranges................................................................0.0 to 199.9

μ

A

200 to 1999

μ

A

2.00 to 10.00 mA

Accuracy

DC to 1 kHz.....................................................

±

(1 % of reading + (1

μ

A or 1 LSD, whichever is greater))

1 to 100 kHz ....................................................

±

(2 % of reading + (1

μ

A or 1 LSD, whichever is greater))

100 kHz to 1 MHz............................................

±

(5 % of reading + (1

μ

A or 1 LSD, whichever is greater))

Mains on applied part test voltage ......................100 %

±

5 % of Mains for AAMI, current limited to 1 mA

±

25 % per AAMI

100

%

±

5 % of Mains for 62353 current limited to 3.5 mA

±

25 % per 62353

100

%

±

5 % of Mains for 60601 current limited to 7.5 mA

±

25 % per 60601

Differential leakage

Ranges................................................................10 to 199

μ

A

200 to 2000

μ

A

2.00 to 20.00 mA

Accuracy..............................................................

±

(10 % of reading + (2 counts or 20

μ

A, whichever is greater))

Insulation resistance

Ranges................................................................0.5 to 20.0 M

Ω

20.0 to 100.0 M

Ω

Accuracy

20 M

Ω

Range..................................................

±

(2 % of reading + 0.2 M

Ω

)

100 M

Ω

Range................................................

±

(7.5 % of reading + 0.2 M

Ω

)

Source test voltage .............................................500 V dc (+20 %, -0 %) 1.5

±

0.25 mA short-circuit current or 250 V dc selectable

Maximum load capacitance.................................1

μ

F