3 descriptions of commands, 1 *cls, 2 *ese <value – Fluke 1595A User Manual

Page 84: 3 *esr, 1 *cls 8.7.3.2, Ese <value

Advertising
background image

1594A/1595A Super-Thermometer

Interface Commands

74

Command

Function

Associated Setting or

Function (screen)

SYST:PASS:GROU:RES

Sets the password Resistor Protection setting

Resistor Protection (Password)

SYST:PASS:GROU:TIME

Sets the password Time/Date Protection setting

Time/Date Protection (Password)

SYST:PASS:NEW

Changes the password

Password (Password)

SYST:PON:MIN?

Reads the time in minutes since power-on

SYST:TIME

Sets the Time of the system clock

Time (Time and Date)

SYST:TIME:DAYL

Sets the Daylight Saving automatic adjust setting

Daylight Saving (Time and Date)

TEST:CURR:CONF

Sets the configuration of the sensing current test

CURRENT TEST (Calibration)

TEST:CURR[:STAT]

Starts or stops the sensing current test

CURRENT TEST (Calibration)

TEST:LIN:ABOR

Cancels a linearity self-test

LINEARITY TEST (Calibration)

TEST:LIN:INIT

Starts linearity calibration

LINEARITY TEST (Calibration)

TEST:LIN:PAR?

Reads the suggested linearity calibration parameters

LINEARITY TEST (Calibration)

TEST:LIN:REP<n>?

Reads the linearity self-test report

LINEARITY TEST (Calibration)

TEST:LIN:STAT?

Reads the state of the linearity self-test

LINEARITY TEST (Calibration)

TEST:LIN:TIME?

Reads the time remaining of the linearity self-test

LINEARITY TEST (Calibration)

TEST:LIN:TOL<n>?

Reads a linearity calibration error tolerance

LINEARITY TEST (Calibration)

TEST:SYST:ABOR

Cancel a system self-test

SYSTEM TEST (Calibration)

TEST:SYST:INIT

Initiate a system self-test

SYSTEM TEST (Calibration)

TEST:SYST:REP<n>?

Reads the report from a system self-test

SYSTEM TEST (Calibration)

TEST:SYST:STAT?

Read the completion state of the system self-test

SYSTEM TEST (Calibration)

UNIT:TEMP

Set the temperature Unit

Unit (Temperature Unit)

8.7.3 Descriptions of Commands

The function of each command as well as an example of each is listed below.

8.7.3.1 *CLS

Clears all status registers and the error queue. Specifically, this command clears the Standard Event Status

Register (see Section 8.7.3.3, *ESR?, on page 74), Operation Status Event Register (see Section 8.7.3.89,

STAT:OPER?, on page 94), Questionable Status Event Register (see Section 8.7.3.92, STAT:QUES?, on

page 94), and system error queue (see Section 8.7.3.116, SYST:ERR?, on page 98). The Status Byte Reg-

ister (see Section 8.7.3.7, *STB?, on page 76) is updated to show the new state of the registers.

8.7.3.2 *ESE <value>

Example command: *ESE 16
Query form: *ESE?
Sets the Standard Event Status Enable Register. This mask register determines which corresponding bits of the

Standard Event Status Register (see Section 8.7.3.3, *ESR?, on page 74) affect the ESB summary message

(bit 5) of the Status Byte Register.

8.7.3.3 *ESR?

Example response: 160
Reads and clears the Standard Event Status Register. This indicates any of several events defined by the IEEE-

488.2 standard and which are listed in the following table. Reading the Standard Event Status Register clears

it. The register is also cleared by the *CLS command.

Advertising
This manual is related to the following products: