Interface Commands

80

Sets the Delta Channel setting (see Section 5.7.4.3, STATISTICS SETTINGS (F3), on page 49).
Parameter <channel> is the channel number to use as the reference, which may be 1 through 4 for the cor-

responding front channel input, 5 through 14 for channels of the first attached scanner, and 15 through 24 for

channels of the second attached scanner.
This setting is only meaningful when Delta Mode is set to CHANNEL.
The *RST command sets Delta Channel to 2.

8.7.3.24 CALC<n>:DELT:DATA?

Example command: CALC3:DELT:DATA?
Example response: 0.104428
Reads a channel’s delta result (see Section 5.7.4.2, FIELD SETTINGS (F2), on page 48).
Suffix <n> is the channel number: 1 through 4 for the corresponding front channel input, 5 through 14 for

channels of the first attached scanner, and 15 through 24 for channels of the second attached scanner.
No result will be available if no measurement has been acquired on a designated Delta Channel or if the two

measurements have different units. If no result is available, no response is returned and error -230,“Data cor-

rupt or stale” is generated in the error queue.

8.7.3.25 CALC:DELT:MODE OFFS|CHAN

Query form: CALC:DELT:MODE?
Alternate parameter: DEF
Sets the Delta Mode setting (see Section 5.7.4.3, STATISTICS SETTINGS (F3), on page 49).
If the parameter is OFFS (OFFSET), the delta result is calculated by subtracting the Delta Offset setting’s

value from the channel’s measurement. If the parameter is CHAN (CHANNEL), the delta result is calculated

by subtracting the measurement from the channel determined by the Delta Channel setting from the channel’s

measurement.
The *RST command sets Delta Mode to the default setting which is OFFS.

8.7.3.26 CALC:DELT:OFFS <value>

Query form: CALC:DELT:OFFS?
Sets the Delta Offset setting (see Section 5.7.4.3, STATISTICS SETTINGS (F3), on page 49).
This setting is only meaningful when Delta Mode is set to OFFSET.
The *RST command sets Delta Offset to the default value which is 0.

8.7.3.27 CALC<n>:TYPE TEMP|RES|RAT

Query form: CALC<n>:TYPE?
Alternate parameter: DEF
Sets the channel’s Calculation setting.
Suffix <n> specifies the channel number: 1 through 4 for the corresponding front channel input, 5 through 14

for channels of the first attached scanner, and 15 through 24 for channels of the second attached scanner. The

default setting may depend on the probe and reference resistor settings.

8.7.3.28 DATA:REC:ALL <boolean>

Query form: DATA:REC:ALL?
Alternate parameter: DEF
Sets the data recording Record Statistics setting (see Section 5.7.3.4.1, RECORDING SETTINGS (F1), on

page 42). The default is ON.