60 inp:prob:test? <id>,<resistance, 61 inp:rear<n>:rs:iden <id>|var|none, Inp:prob:test? <id>,<resistance – Fluke 1595A User Manual

Page 97: Inp:rear<n>:rs:iden <id>|var|none

Advertising
background image

87

Remote Operation

Interface Commands

C0 – C7

C0 – C7

exponential number

Thermistor type settings
Maximum Temperature

MAX_TEMP

decimal number, dependent on Unit setting

Conversion

CONV

R_POLY|T_POLY|NONE

R(T) polynomial conversion settings
B0 – B3

B0 – B3

exponential number

T(R) polynomial conversion settings
A0 – A3

A0 – A3

exponential number

The settings that are applicable to the probe definition depends on the Type setting and Conversion setting, and

may depend on other settings also, such as an ITS-90 subrange.
Any channel that has the probe assigned to it and has Calculation setting set to temperature will have Calcula-

tion (see Section 8.7.3.27, CALC<n>:TYPE TEMP|RES|RAT, on page 80) changed to resistance when the

probe’s Type setting is set to resistor or the Conversion setting is set to none.

8.7.3.60 INP:PROB:TEST? <id>,<resistance>

Example command: INP:PROB:TEST? “PRT_A46002”,65.449411
Example response: 419.527,C
Reads the temperature calculated from resistance using the probe definition.
Parameter <id> is the ID of a probe definition in the Probe Library. If <id> does not match the ID of a probe

definition in the Probe Library, error -224,“Illegal parameter value” is generated in the error queue.
Parameter <resistance> is the resistance from which temperature is calculated.
The normal response contains the calculated temperature and the unit, separated by a comma:
<Temperature>,<unit>
The <temperature> value is converted to units determined by the Unit setting (see Section 5.7.4.5, TEMPERA-

TURE UNIT (F5), on page 50). If the probe Type is RESISTOR or Conversion is NONE, the <resistance>

parameter value is returned instead, with <unit> “O” for ohms. If a valid response cannot be returned because

the given resistance is out of range or the probe definition is invalid, no response is given and error -230,“Data

corrupt or stale” is generated in the error queue.

8.7.3.61 INP:REAR<n>:RS:IDEN <id>|VAR|NONE

Example command: INP:REAR1:RS:IDEN “R25_01322”
Example command: INP:REAR2:RS:IDEN NONE
Query form: INP:REAR<n>:RS:IDEN?
Example query response: “R25_01322”
Example query response: VAR
Sets the resistor definition assigned to a rear reference resistor input. (Use the command INP<n>:RS:IDEN to

assign a resistor definition to a front channel input.)
Suffix <n> specifies the rear Rs input number and must be either 1 or 2.
Parameter <id> must be either the ID of a resistor definition in the Resistor Library, “VAR” (variable), or

“NONE”. Enclosing quotations marks are optional. If <id> does not match the ID of a resistor definition in the

Resistor Library or an allowed option, error -224,“Illegal parameter value” is generated in the error queue.
If a valid resistor ID is assigned to the input, the response to the query form is the ID of the resistor enclosed

in quotation marks. If the resistor assignment is variable or none, the response is “VAR” or “NONE” respec-

tively, without quotation marks.