Environmental conditions – Fluke 43B User Manual

Page 48

Advertising
background image

Fluke 43B

Users Manual

44

Environmental Conditions

Environmental ........................................MIL 28800E, Type 3, Class III, Style B

Temperature

During operation ....................................................... 0 to 50

°

C (32 to 122

°

F)

While stored ...........................................................-20 to 60

°

C (-4 to 140

°

F)

Humidity

During operation:

0 to 10

°

C (32 to 50

°

F)................................................... non-condensing

10 to 30

°

C (50 to 86

°

F)......................................................... 95 % ± 5 %

30 to 40

°

C (86 to 104

°

F).........................................................75 % ± 5 %

40 to 50

°

C (104 to 122

°

F).......................................................45 % ± 5 %

While stored:

-20 to 60

°

C (-4 to 140

°

F) ................................................ non-condensing

Altitude

During operation .............................................................. 4.5 km (15 000 feet)

The maximum input and floating voltage is 600 V rms up to 2 km.
Linearly derating from 600 down to 400 V rms between 2 km to 4.5 km.

While stored ..................................................................... 12 km (40 000 feet)

Vibration................................................................................................. max. 3g

Shock ................................................................................................... max. 30g

Electromagnetic Compatibility (EMC)

Emission...........................................................................EN 50081-1 (1992):

EN55022 and EN60555-2

Immunity...........................................................................EN 50082-2 (1992):

IEC1000-4-2, -3, -4, -5

(See also Tables 1 to 3)

Enclosure Protection ..............................................................IP51, ref: IEC529