Specifications

Electromagnetic Immunity

3

45

Electromagnetic Immunity

The Fluke 43B, including standard accessories, conforms with the EEC
directive 89/336 for EMC immunity, as defined by IEC1000-4-3, with the
addition of the following tables.

Disturbance with test leads TL24 or Current Clamp 80i-500s

Volts / amps / hertz

Resistance,

Capacitance

Power

Harmonics

Table 1

No visible disturbance

E = 3 V/m

E = 10 V/m

Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 1 GHz

(-)
(-)

(-)
(-)

(-): no visible disturbance

Disturbance with test leads TL24 in scope mode

V AC+DC (True RMS)

Table 2

Disturbance less than 1 %
of full scale

E = 3 V/m

E = 10 V/m

Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 200 MHz
Frequency: 200 MHz - 1 GHz

2 V/div - 500 V/div

500 mV/div - 500 V/div

(-)

10 V/div - 500 V/div

2 V/div - 500 V/div

5 mV/div - 500 V/div

(-): no visible disturbance

Table 3

Disturbance less than 10 % of
full scale

E = 3 V/m

E = 10 V/m

Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 200 MHz
Frequency: 200 MHz - 1 GHz

1 V/div

200 mV/div

(-)

5 V/div
1 V/div

(-)

(-): no visible disturbance

Ranges not specified in Tables 2 and 3 may have a disturbance of more than
10 % of full scale.