GE E4H-CE Series User Manual

Page 34

Advertising
background image

29

TROUBLESHOOTING

SYMPTOM

Declining rejection (high

permeate conductivity)

(continued)

Machine not operating

POSSIBLE CAUSES

O-ring seal broken or dam-

aged

Change in incoming water

quality

Inaccurate conductivity moni-

tor or fouled probe

Pressurized storage switch or

float switch has cut power to

machine

External control or permeate

purge control contacts have

shut machine down

REMEDIES

Replace O-ring, check the

sealing surfaces on the

O-ring groove, intercon-

nectors and end caps.

Replace damaged parts.

Open the concentrate valve

and flush. Test the water

for pH, hardness, TDS, and

iron content. A water

analysis should be sent to

GE Osmonics for review.

Calibrate the monitor with

a DS standard solution or

check the readings with

another conductivity meter.
Replace or clean the probe.

Check the connections
between the probe and

monitor.

Check the permeate back

pressure or position of float

in the storage tank.

Check for proper orienta-

tion of control contacts.

Restart machine manually

or wait for automatic start-

up.