The following guide is for reference purposes only. Contact a Metrologic
representative at 1-800-ID-METRO or 1-800-436-3876 to preserve the
limited warranty terms.

MS7220 Series Troubleshooting Guide

SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

All Interfaces

No LEDs, beep
or motor spin

No power is being
supplied to the
scanner

Check transformer, outlet
and power strip. Make sure
the cable is plugged into the
scanner

No LEDs, beep

No power is being
supplied to the
scanner from host

Some host systems cannot
supply enough current to
power ArgusScan™. Use
the power supply included
with the scanner.

3 beeps on
power up

Non-volatile RAM
failure

Contact a Metrologic
Representative, if the unit
will not hold the
programmed configuration

Continuous
razz tone on
power up

RAM or ROM failure

Contact a Metrologic
Representative, if the unit
will not function

Razz tone and
green LED flash
at power up

VLD failure

Contact a Metrologic
Representative

Razz tone and
both LEDs flash
at power up

Scanner motor failure

Contact a Metrologic
Representative

Multiple scans
upon
presentation of
code

Same symbol timeout
set too short

Adjust same symbol timeout
for a longer time

The unit powers
up, but does
not scan and/or
beep

Beeper disabled. No
tone selected

Enable beeper. Select tone