Electric connection

The connection needs to be made in the wiring box at the back. Remove the
coverplate (G) in order to reach the connection clamps.

The connection cable needs to be brought in through the swivel (H) on the side of
the wiring box.

It is necessary to ground the dryer for your personal safety and to ensure a good
operation.

115 and 208-240 V - 1 ph

The mains wires (Ll) and (L2) should be connected to the 2 left clamps and the
yellow/green grounding wire (PE) should be connected to the grounding clamp.

208-240 and 480 V - 3 ph

The 3 mains wires (L1), (L2) and (L3) should be connected to the 3 left clamps and
the yellow/green grounding wire (PE) should be connected to the grounding
clamp.

380-415V - 3 ph

The 3 mains wires (L1), (L2) and (L3) should be connected to the 3 left clamps,
the blue neutral (N) should be connected to the right clamp and the yellow/green
grounding wire (PE)
should be connected to the grounding clamp.

G

H

Dryers must be electrically grounded by a separate #14 or larger green wire from the grounding terminal
within the Service Connection Box, to a cold water pipe. In all cases, the grounding method must comply
with local electrical code requirements; or in the absence of local codes, with the National Electrical Code,
ANSI/NFPA 70 or the Canadian Electrical Code, CA C22.1.

See wiring diagram furnished with dryer. Do not change wiring without consulting the factory, as you may
void the factory warranty. DO NOT CONNECT THE DRYER TO ANY VOLTAGE OR CURRENT
OTHER THAN THAT SPECIFIED ON THE DRYER RATING PLATE. (Wiring diagram is located on rear
wall of dryer.)

All panels must be in position before operation of dryer.

115 and 208-240 V - 1 ph

208-240 and 480 V - 3 ph

380-415V - 3 ph

N

L3

L2

L1