SECTION 4—CHECKLIST AND CYLINDER INSPECTION

Part No 1100873

21

HomeFill

®

Compressor

FIGURE 4.1 Cylinder Prefill Inspection

Improper filling or abuse of this cylinder, or failure to follow these safety
instructions may cause serious bodily injury or death. High pressure
oxidizing gas vigorously accelerates combustion.

DO NOT alter or modify cylinder or related components.
DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment
and connections free of oil and grease to avoid violent ignition resulting
in serious bodily injury or death.

For medical applications. Use only as directed by a licensed
practitioner. Uninterrupted use of high concentrations of oxygen over
a long duration, without monitoring its effect on oxygen content of
arterial blood, may be harmful.

Oxygen Cylinders filled by the Invacare HomeFill oxygen
system shall be used for personal use only.

"

Not to be

filled for resale or use by professional users.

"

This cylinder is to be used with the Invacare HomeFill
oxygen system. This cylinder contains 93 ± 3% oxygen
produced by an oxygen concentrator using the pressure swing
adsorption process.

DO NOT handle cylinder or use contents until you are professionally
trained, including emergency procedures.

Use in accordance with Invacare HomeFill oxygen system Operator’s
Manual. Open valve slowly. Close valve after each use and when
empty. Secure cylinder during storage and use. No smoking in cylinder
area. Keep away from heat, flame and spark. Keep out of reach of
children. DO NOT drop.
Filling of this gas is performed by Invacare HomeFill oxygen system
Only.
Cylinders that have been refinished and/or exposed to elevated
temperatures must be hydrostatically tested before filling. Cylinder
temperature should not exceed 125°F (52°C). Aluminum cylinders
subject to the action of fire or heated to temperatures in excess of
350°F must be withdrawn from service and condemned by trained
personnel. DO NOT use caustic paint strippers. Valve and safety
relief devices must be removed and replaced by trained and
authorized personnel. DO NOT alter or change valve assembly.

DO NOT REMOVE THIS PRODUCT LABEL

Not for Commercial Resale

Part Number 1125090 Rev D

Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford, FL 32773
1-800-832-4707

HomeFill

Cylinder

Label