SECTION 6—INDICATOR LIGHTS

HomeFill

®

Compressor

30

Part No 1100873

SECTION 6—INDICATOR LIGHTS

INDICATOR LIGHT

COLOR

COMPRESSOR

STATUS

EFFECT

None

Compressor is Off.

O

2

flowing to user if

concentrator is on.

O

2

BELOW NORMAL -

YELLOW

Compressor is On and
warming up
(0-3 minutes after
power is On).

Cylinder filling not
started.
O

2

flowing to user if

concentrator is On.

O

2

from concentrator is

< 90% (any time after
three minute warmup).

Cylinder filling stops, will
restart when O

2

> 90%.

O

2

flowing to user if

concentrator is On.

FILLING - GREEN

Compressor is On and
cylinder is filling. O

2

from concentrator is
>90% (any time after
three minute warmup).

Cylinder is filling.

O

2

flowing to user if

concentrator is On.

FULL - GREEN

Compressor is On. Cyl-
inder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not attached
properly - check con-
nection (FULL light is
On in <10 mins.).

Cylinder filling stops.
Cannot fill another
cylinder until
compressor is turned Off
and On again.

O

2

flowing to user if

concentrator is On.

ATTENTION - RED
(Audible Alarm)

Compressor is On.
Cylinder NOT filling
Causes:
Full cylinder attached
when unit turned On -
replace with empty
cylinder
or
Internal failure -
contact dealer.

Compressor does not
begin to fill or stops
filling.
Unit must be turned Off
and On before
filling can occur.

O

2

flowing to user if

concentrator is On.

Advertising