SECTION 5—COMPRESSOR OPERATION

HomeFill

®

Compressor

28

Part No 1100873

Turning the Compressor On

NOTE: For this procedure, refer to FIGURE 5.5 on page 29.

1. Make sure the concentrator is On. Refer to the concentrator 

Owner’s Manual.

CAUTION

The concentrator flow rate to the patient MUST be set at 
maximum: 3 L/min or less for Platinum 5, XL, and Perfecto

2

 and 

5 L/min for Platinum 10. Otherwise, the oxygen level to the 
compressor will be <90% and the compressor will not fill the 
cylinder.

2. Set the concentrator flow rate to the patient to 3 L/min or less 

for Platinum 5, XL, and Perfecto2 and 5 L/min for Platinum 
10. Refer to the concentrator Owner’s Manual.

3. Make sure the cylinder is connected to the compressor. Refer 

to Connecting the Cylinder to the Compressor on page 23.

4. Push the power switch on the control panel to the On ( ) 

position.

5. Examine the control panel. The following sequence of events 

should occur:

A. 0‐3 minutes ‐ The O

2

 Below Normal (YELLOW) light will 

be lit while the compressor warms up.

B. After three minutes ‐ The FILLING (GREEN) light will be 

on while the cylinder is filling.

C. The FULL (GREEN) light will be on when the cylinder is 

finished filling. Proceed to STEP 6.

6. Remove the full cylinder. Refer to Disconnecting the 

Cylinder from the Compressor on page 26.

7. Push the power switch to the Off (O) position.