SECTION 5—COMPRESSOR OPERATION

Part No 1100873

27

HomeFill

®

Compressor

ƽ WARNING

The fillport cover on the connector fillport and the cylinder 
fillport MUST be replaced after filling and whenever not in use. 
If either of the fillport covers are missing contact your healthcare 
provider before using.

6. Place the fillport covers onto the connector fillport and 

cylinder fillport.

NOTE: When using the post valve type cylinder, attach an oxygen 
regulator per manufacturers instructions.

7. Examine the cylinder gauge to ensure cylinder is full (needle 

pointing into GREEN area). If the cylinder is not full, perform 
the following:

A. Repeat Connecting the Cylinder to the Compressor on 

page 23.

B. If cylinder is not full after second attempt to fill, contact 

the provider.

FIGURE 5.4 Fillport Covers

Cylinder

Fillport

Cover

Connector

Fillport Cover