SECTION 4—CLEANING AND DISINFECTING

Stratos™Portable Plus

28

Part No 1150755

2. Disassemble nebulizer chamber by turning 

counterclockwise and separating.

3. Fill two plastic containers or bowls:

A. Washing: Hot water/detergent solution.

B. Soaking: Hot water/vinegar solution (one part vinegar 

to three parts water).

4. Thoroughly clean the nebulizer and mouthpiece in hot 

water/detergent solution.

5. Remove from the water/detergent solution and rinse with 

clear hot tap water.

6. Soak in hot water/vinegar solution for 30 minutes.

7. Rinse with hot tap water again and air dry thoroughly.

NOTE: If using medical disinfectant cleaners, follow 
manufacturer’s instructions carefully.

8. Keep the outer surface of the tubing dust‐free by wiping 

regularly.

The nebulizer air tubing does not have to be washed internally 
because only filtered air passes through the tubing.