Problem solution – Invacare IRC 1730 User Manual

Page 34

Advertising
background image

SECTION 6—TROUBLESHOOTING

Stratos™Portable Plus

34

Part No 1150755

Nebulizer does not work,
and the light indicator is red.

The thermal protector has been activated due
to:
a. The device operating outside the conditions
outlined in this manual.
b. The device has been operating near heat
sources or in hot environments.
Allow the nebulizer to cool down for a
minimum of 40 minutes. Wait until the RED
light turns off before turning the nebulizer on
again.

Optional battery is not
working optimally.

Battery was not properly conditioned (charged
for at least 6 hours before first use).
If it has been in use for approximately four
months, it may be time to recondition the
battery.

Optional battery is not
charging.

Battery will not charge if connected to the
nebulizer.
Check all power cord connections and cord
conditions.
Check power outlet.

PROBLEM

SOLUTION

Advertising