SECTION 4—CLEANING AND DISINFECTING

Part No 1150755

29

Stratos™Portable Plus

Cleaning the Mouthpiece

1. Boil the mouthpiece in water for a maximum of five 

minutes or autoclave at a maximum of 121° C for ten 
minutes.

2. Allow the mouthpiece to air dry.
3. Once dry, the mouthpiece may be stored inside the 

nebulizer top.

Changing the Nebulizer

The nebulizer is designed for single patient use.

The nebulizer MUST be replaced after a long period of 
inactivity, or when it is obstructed by dry medication, dust, 
etc. The reusable nebulizer should last from six months to one 
year with general use. The disposable nebulizer should last 30 
days.

Cleaning the Compressor

DANGER

DO NOT remove outer case of this unit. Removing the

case creates a serious electrical shock hazard. All

disassembly and maintenance of this unit MUST be done

by a qualified technician.