31

Number of

flashes

Fault

Impact on

Scooter

Notes

4

Current Limit
Time Out

Drive inhibited

The scooter has drawn too much current
for too long, possibly because the motor
has been over-worked, jammed or
stalled. Turn the scooter power off, leave
it off for a few minutes, and then turn the
power back on again.

The controller has detected a short-
circuited motor. Check the cable harness
for short and check the motor: Contact
your authorised Invacare® Dealer.

5

Brake Fault

Drive inhibited

Check that the declutching lever is in the
engaged position.

The park brake coil or wiring is faulty.
Check the park brake and wiring for open
or short circuits. Contact your authorised
Invacare® Dealer.

6

Out Of Neutral
At Power Up

Drive inhibited

Throttle is not in neutral position when
turning key switch on. Return throttle to
neutral, turn power off, and back on
again.

Throttle may need to be re-calibrated.
Contact your authorised Invacare®
Dealer.

7

Speed Pot
Error

Drive inhibited

The throttle or its wiring may be faulty or
incorrectly set up. Contact your
authorised Invacare® Dealer.

8

Motor Volts
Error

Drive inhibited

The motor or its wiring is faulty. Contact
your authorised Invacare® Dealer.