8.2.3.1 Removing the batteries

Requirements:

Wrench 11 mm

Remove the seat.

Remove the battery and motor compartment cover.

Disconnect the main cable harness.

Open the belts that hold the batteries in place (1).

Loosen battery clamp of the blue cable at the negative pole
of the battery with the wrench and remove cable.

Loosen the battery clamp of the red cable at the positive
pole of the battery with the wrench and remove cable.

Repeat procedure on the other side.

NOTE
Fitting of new batteries is done in reverse order.