8.2.3.1 Removing the batteries
Requirements:
•
Wrench 11 mm
•
Remove the seat.
•
Remove the battery and motor compartment cover.
•
Disconnect the main cable harness.
•
Open the belts that hold the batteries in place (1).
•
Loosen battery clamp of the blue cable at the negative pole
of the battery with the wrench and remove cable.
•
Loosen the battery clamp of the red cable at the positive
pole of the battery with the wrench and remove cable.
•
Repeat procedure on the other side.
NOTE
Fitting of new batteries is done in reverse order.