SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B User Manual

Page 10

Advertising
background image

3-3

2.4 Electrical

Connection


The system utilizes Universal Switching Power Supply, and will
accept voltages from 90 – 260 VAC, 50-60 Hz.

WARNING: Do not bypass the safety ground connection as
a serious shock hazard could result.

2.5 Solvent

Preparation

Proper solvent preparation will prevent a great number of pumping

problems. The most common problem is bubble formation, which may
affect the flow rate consistency. Aside from leaky fittings, the problem
of bubble formation arises from two sources: solvent out-gassing and
cavitation. Filtration of HPLC solvents is also required.

2.5.1 Solvent

Out-gassing and Sparging

Solvent out-gassing occurs because the mobile phase contains

dissolved atmospheric gases, primarily N2 and O2. These dissolved
gases may lead to bubble formation and should be removed by
degassing the mobile phase before or during use. The best practical
technique for degassing is to sparge the solvent with standard
laboratory grade (99.9+%) helium. Helium is only sparingly soluble
in HPLC solvents, so other gases dissolved in the solvent diffuse into
the helium bubbles and are swept from the system. Solvent filtration is
not an effective alternative to helium degassing.

It is recommended that you sparge the solvent vigorously for 10 to

15 minutes before using it. Then maintain a trickle sparge during use
to keep atmospheric gases from dissolving back into the mobile phase.
The sparged solvent must be continually blanketed with helium at 2 to
3 psi. Non- blanketed, sparged solvents will allow atmospheric gases
to dissolve back into the mobile phase within four hours.

Solvent mixtures using water and organic solvents (like methanol or
acetonitrile) hold less dissolved gas than pure solvents. Sparging to
reduce the amount of dissolved gas is therefore particularly important
when utilizing solvent mixture.

Even with sparging some out-gassing may be occur. A back

pressure regulator installed after the detector flow cell will help
prevent bubbles from forming and thus limit baseline noise.

WARNING: Always release pressure from the pump slowly. A rapid
pressure release could cause the pulse damper diaphragm to rupture.

Advertising