Appendix a – SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B User Manual

Page 39

Advertising
background image

A-1

APPENDIX A

A.1

Rear Panel Serial Communications Port - Pump


An RS-232C modular jack is provided on the back panel. A

computer, with appropriate software, can be used as a remote
controlling device for pump operation via this connection.

A.1.1 Hardware Implementation

The REMOTE INPUT serial communications port is configured

for 9600 baud, 8 data bits, 1 stop bit, and no parity. The connector is a
standard RJ-11 modular telephone type jack. When looking at the
connector on the rear panel of the pump, pin 1 is at the top and pin 6 is
at the bottom. The pin-out is:

Pin

Function

1,

6

Ground

2

DSR (Handshaking input to pump)

3

RXD (Serial data input to pump)

4

TXD (Serial data output from pump)

5

DTR (Handshaking output from pump)


Special wiring considerations: Use the following chart for

interfacing the pump's serial communications port to either a 25-pin or
a 9-pin COM port on an IBM-PC type computer.

Pump (RJ11)

Signal

IBM (DB25)

a

IBM (DB9)

b

1,

6

Ground

7

5

2

DSR

20

4

3

RXD

2

3

4

TXD

3

2

5

DTR

6

6

a

Jumper pins 4, 5, and 8 on DB25.

b

Jumper pins 1, 7, and 8 on DB9.

Part Description

Part Number

Modular

Cable

12-0677

Adapter RJ-11 to DB9

12-0672

Adapter RJ-11 to DB-25

12-0671

A.1.2 Hand-Shaking

The pump uses hardware handshaking. The pump will not

transmit on the TXD output if the DSR input is at a low logic level.
And, the pump will not receive on the RXD input when the DTR

Advertising