SP-515

English

3

DEVILBISS 4-LITER SERIES

DEVILBISS 5-LITER SERIES

Catalog Number

515UK

515DS, 515DZ

515KS, 515KZ, 515NS

Delivery Rate (Lower delivery rates

1 to 4 LPM

1 to 5 LPM

available for low flow applications)

Maximum Recommended Flow

4 LPM

5 LPM

Outlet Pressure

8.5 psig (58.6 kPa)

Electrical Rating

230 V, 50 Hz, 1.4 Amp

115 V, 60 Hz, 4.2 Amp

220-230 V~, 50 Hz, 1.7 Amp

230 V~, 60 Hz, 1.9 Amp

Operating Voltage Range

195-253 V~, 50 Hz

97-127 V~, 60 Hz

187-255 V~, 50 Hz
218-253 V~, 60 Hz

Oxygen Percentage

1-4 LPM = 93%±3%

1-5 LPM=93%±3%

Operating Altitude

(tested at 21°C only)

Across the voltage range:

Across the voltage range:

Across the voltage range:

0-1500 M (0-4921 ft)

No degradation of performance

No degradation of performance

No degradation in performance

Tested at nominal voltage only:

Tested at 230V/50Hz only:

1500-3000 M (4921-9842 ft)

Not recommended/Not tested

No degradation in performance

~4% oxygen perfomance drop at 5 LPM

3000-4000 M (9842-13123 ft)

Not recommended/ Not tested

Not recommended/Not tested

Not recommended/Not tested

Operating Environment Range*

No degradation in performance across

No degradation in performance across

~4% oxygen performance drop at 5 LPM

10°C to 35°C, humidity range of 30% to 75%

the operating voltage range

the operating voltage range

and 35°C condition. No degradation in

(tested at sea level).

(tested at sea level).

performance across the operating voltage

range at other liter flows (tested at 670 M).

Power Consumption

295 Watts Average

400 Watts Average

230v / 50 Hz - 365 Watts Average
230v / 60 Hz - 435 Watts Average

Weight

53 lbs. (24.5 Kilograms)

52 lbs. (23.5 Kilograms)

53 lbs. (24.5 Kilograms)

Sound Level (ISO 8359:1996 from front)

50.5 dbA overall average

52.5 dbA overall average**

50.5 dbA overall average (50Hz)

Dimensions

27.75"H x 16"W x 14"D (70.5 x 40.6 x 35.6 cm)

Pressure Relief Valve

44 psig±3psig (303 kPa±21 kPa)

Operating System

Time Cycle / Pressure Swing

For units equipped with an OSD,

Units with serial numbers less than H20000: 85%±2% (The audible alarm will alert at approximately 75%)

the visible “low oxygen” indicator

Units with serial numbers of H20000 and greater: 83.5%±1.5%

will activate at the following level

(The audible alarm will alert at approximately 75%. At less than 60%, the red “service required” light will activate.)

Storage Conditions

-40°C to 70°C, humidity range of 10% to 100%, including condensation

Equipment Class and Type

Class II Equipment Double Insulated Type B Applied Part

Approval Body

TUV

CSA

TUV

and

IEC 601-1+A1+A2

CAN/CSA-C22.2 No. 601.1-M90

IEC 601-1+A1+A2

Safety Standard

EN 60601+A1+A2

EN 60601+A1+A2

ISO 8359:1996

ISO 8359:1996

(515KZ excluded)

EMC Compliance To:

IEC 601-1-2

*Note: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515DS verified at 670m.
**Note: This corresponds with a sound level of 50 dbA overall average as measured according to the guidelines of the now defunct ANSI Z79.13-1981 standard.
Specifications subject to change without notice.