4 reagent status, Reagent status, Operation – LEICA ASP300 User Manual

Page 52

Advertising
background image

52

Instruction Manual V 1.3 – 12/2003

5.4 Reagent status

In the reagent status screen (

Fig. 32

) you can display and update the

actual reagent status of each station and you can set the stations
as full or empty.

Fig. 32

Warning thresholds that trigger
warning messages appear on a
red background.

To change a reagent status,
highlight the corresponding line
and press the corresponding
key symbol .

5.

Operation

Reagent warnings

If the RMS is enabled, the instrument will give
you warning messages when reagents have
been used for too long.

Warning messages for normal reagents are
given based on up to 3 criteria; warning mes-
sages for paraffin wax on up to 6 criteria.

The warning messages are displayed at the
end of a cleaning cycle. From there, you can
go directly to the REAGENT STATUS screen.

In addition, warning messages (on a yellow
background) are displayed in the program
start screens.

Criteria for warning messages

For normal reagents, the following warning cri-
teria apply:

number of blocks processed

number of programs carried out

number of day lapsed

since the reagent has last been changed.

For paraffin wax, three additional criteria can
be applied:

number of blocks processed

number of programs carried out

number of days lapsed

since the wax has last been cleaned.