5.5 System Monitor

The SYSTEM MONITOR screen supplies important infor-
mation on the Leica ASP300 system status.

After replacing the filter press RESET.

ROTARY VALVE RETRIES:

If a figure higher than 50 is indicated, a preventive maintenance is
recommended.

FILL/DRAIN:

This screen displays the average fill / drain times of all stations
(in seconds).

Short term = average duration of last 5 fills / drains.
Long term = average duration of last 20 fills / drains.

If one of the figures displayed exceeds 250, call Leica Technical
Service.

PUMP AGE:

Indication of hours of service of the pressure/
vacuum pump. After approx. 1,000 hours of
use, have Leica Technical Service carry out a
preventive maintenance of the pump.

CARBON FILTER AGE:

Indication of hours of service of the activated
carbon filter. The carbon filter must be ex-
changed after approx. 4 months of use.

Fig. 33