MAINTENANCE

44

N o t e :

If you disinfected the meter

with a 10% bleach solution, you may
not have completely removed the
bleach residue from the gray area
inside the test strip holder cover. Wipe
this area with a cloth dampened with
water and dry thoroughly. Failure to
remove the excess bleach may result in
an error message (

,

) or an

inaccurately high result.

N o t e :

Clean the outside of the

meter by gently wiping it with a cloth
dampened with water. Dry it with a
soft cloth.

6.

If the

Er 5 symbol

appears after cleaning,
check to see if you
inserted the test strip
holder completely and
securely (see step 4 in
this section).