Component description, Installation – Precision Trading PM5200 User Manual

Page 10

Advertising
background image

9

Blender

Air-Oxygen

COMPONENT dESCRiPTiON

ITEM

DESCRIPTION

F

Air inlet Fitting

A male DISS or NIST air fitting with one way valve

that is used to connect an air supply hose

.

g

Alarm

An audible alarm that sounds due to an excessive

pressure drop or deletion of either gas supply.

H

Manifold Outlet (Optional)

manifold with 3 primary outlets.

i

Rear Slide Mount with dove tail.

iNSTALLATiON

WARNiNg

Read this User manual before installing or operating the
Air-Oxygen Blender.
Confirm the concentration of air/oxygen with an Oxygen
Analyzer/monitor.

1. Secure the Air-Oxygen Blender to a wall or pole bracket in an

upright position.

2. It is recommended to install a condensation trap in the air

supply line.

3. Connect the air and oxygen supply lines to the appropriate

inlet fittings on the bottom of the Blender.

4. Attach a flowmeter, or other metering device to one of the

outlet ports.

Primary Outlets Flow capacity:

• High Flow Blender (Pm 5200 model) 15 lpm to 120 lpm

• Low Flow Blender (Pm 5300 model) 3 lpm to 30 lpm

Auxiliary Outlet:

This will bleed off some of the air/oxygen mixture to

maintain concentration accuracy at the Low Flow setting.

High Flow Blender (Pm 5200 model) 15 lpm or less

• Low Flow Blender (Pm 5300 model) 3 lpm or less

5. Attach a supply line to the outlet port of the flowmeter.

Advertising
This manual is related to the following products: