Blender

Air-Oxygen

SPECiFiCATiONS

continued

dimensions:

(without fittings)

Depth:

4.9 in

(12.5 cm)

Width:

2.3 in

(5.7 cm)

Height:

4.1 in

(10.4 cm)

Weight:

2.9 lbs

(1.3 kg)

Shipping Weight:

3.5 lbs

(1.6 kg)

Operating Temperature Range:

59°F to 104°F (15°C to 40°C)

Transport / Storage Requirements

Temperature Range:

-10°F to 140°F (-23°C to 60°C)

Humidity:

max 95% Noncondensing

FiO

2

Accuracy:*

± 3% of full scale @ 50 psi (3.4 bar)
± 4% of full scale @ 60 psi (4.14 bar)

Pressure drop:

Low Flow:

≤ 2 psi (0.14 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 10 lpm flow rate at 60% FiO

2

.

High Flow:

≤ 3 psi (0.21 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 30 lpm flow rate at 60% FiO

2

.

The Air-Oxygen Blender has been degreased for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 7767 to meet CE requirement.

dryness and Composition for inlet gases:

Air:

medical Air supply should meet the requirements of

ANSI Z86.1 - 1973 commodity specification for Air, type

1 grade D or better.

Oxygen:

Oxygen supply must meet all requirements of USP

medical Oxygen Grade N.

dew Point:

(ONLY for CE

requirements)

Both inlets should remain 10°F (5.5°C) or more

below the lowest temperature to which the air

distribution system equipment is exposed. At a

temperature of 25°F (-3.9°C) and a pressure of

90 psi (6.33 kg/cm

2

) this equates to 2000 mg/m

3

.

* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the high flow

Blender, and 3 lpm for the low flow Blender.

Specifications are subject to change without prior notice.